    MAS-1

    The Simple Way of Invasive Measurements

    The MAS-1 Probe is an invasive probe for the measurement of mA (tube current) and mAs. Together with Piranha or Cobia Sense  direct reading of mA and mAs as well as waveform are obtained. The MAS-1 Probe connects to the mAs socket of the X-ray generator and to the multimeter. The MAS-1 Probe can be used to measure tube current for both fluoroscopic and radiographic exposures. The probe has a built-in low-pass filter with user-selectable bandwidth of 5 or 500 Hz and accepts a floating input range of ±90 V relative to ground.

    MAS-1 Probe with Piranha & Cobia Sense

    The MAS-1 Probe supports our automatic detector identification system. When connected, the Piranha or Cobia automatically identifies the probe and makes all necessary adjustments without any need for interaction from the user.
    The system also gives full interchangeability of probes between different Piranha systems or Cobia systems, using the same type of probes. 

    Specifications with Black Piranha & Cobia Sense

    Type MAS-1B
    Current range  0.1–3000 mA
    Inaccuracy ±1 % or ±0.01 mA
    Charge range 0.001 mAs-9999 mAs
    Frequency range DC–500 Hz (-3 dB)
    Size 60 x 35 x 72 mm
    Weight 125 g
    Battery life around 100 h
    Rechargeable battery with mini-USB connector
    Cable length 4 m (extension available)
    Connector Hirose

     

     

     

    Article Nr (with Piranha): 9730005-00
    Article NR (with Barracuda): 9730009-00

     