Easy & Fast X-ray Quality Control
Fast for Real
Everything happens in a flash! The Black Piranha is quick, Ocean software is quick and therefore your daily QA-work will go quickly. Ocean can perform instant real-time analysis during your measurements. Ocean also prepares a report in the background as you go. So when the work is done, if you want a complete report of your work – just press print!
Use your tablet/laptop as both an interactive display during the measurements and as a powerful analysis tool when you are back at the office. All your measurements, along with the result and the report will be stored inside your tablet or laptop. No unnecessary, time-consuming data-transfer at the end of the day.
Simply Plug n Play
Everything happens automatically – just plug n play. The Quick Check can almost feel what you are about to do. The Quick Check feature identifies your probe and selects the optimum Piranha settings for your measurements. It can also be programmed to always start with the kind of set up that you usually start your QA-work with. There is really not much you have to think about before you begin – sit back and let the Piranha do the work!
A Piranha Never Forgets
During the measurements you will only see the desired values on your tablet or laptop but since the Piranha continues to work behind the scenes, you always have access to all of the data from each measurement. This enables you to perform for instance trend analysis and study the history of each X-ray equipment – even months after you have completed your measurements.
A Complete QA Solution Instantly
The Piranha comes ready-to-use with everything you need included. It is really an all-in-one multifunction meter. Connect to your computer wireless or via USB and you have a complete QA-system! The number of Piranha models available ensures that you only pay for what you need. As your needs grow, you have unlimited ability to upgrade your Piranha.
Position Check
Piranha has a unique feature that makes it possible to check the position of the detector before measuring. By using the “Position Check” function, you can verify that the detector area is fully irradiated. Possible field inhomogeneities are also neutralized.
RTI Dose Probe
The RTI Dose Probe is an external dose probe, designed to perform very low dose rate measurements for instance on image intensifiers. To avoid or minimize interference with AEC (Automatic Exposure Control) on X-ray equipment, the probe is very small. This also enables to fit it into the table bucky. Since it is a solid-state detector, no corrections for pressure or temperature are needed. Neither is bias voltage. With the extra ordinary sensitivity of the external Dose Probe it is even suitable for scatter and leakage measurements.
Piranha Key Features
All-in-one multifunction X-ray meter | Automatic recognition of external probes |
One-shot HVL for Mammography, Radiography, CT, and Dental | Small, compact & robust – easy to place |
Solid-state detectors = no need to compensate for temperature & pressure | Backscatter protected |
Can measure on scanning beams as well as tomosynthesis | Long-lasting rechargeable battery |
Optimized for X-ray equipment from a large number of manufacturers | Lead-free |
Built-in energy compensation | Always free firmware upgrades |
Can be used together with ion chambers | Firmware is easily updated via Internet or CD |
Wide-range detection of total filtration | Upgradable to any model you want |
100 meters Bluetooth range | Up to 10-year warranty |
Unique detector design to minimize position and rotation dependence | 2-year Calibration Cycle |
Download the Black Piranha folder in various languages:
Size 2.75 MB
Specifications
Size | 133 x 75 x 26 mm |
Weight | Approx. 400 g |
Display | PC or Windows Tablet |
Software | Ocean (Complete Software for QA) |
Databas Support | Yes |
Trend Analysis | Yes |
Interface Type | Built-in bluetooth and USB |
Detector Position Check | Yes |
Simultaneous Dose Measurements | Yes (max 2 detectors) |
Use of Dose Detecor with Active AEC | Yes**** |
Bluetooth Range | 100 m (free-in-air) |
USB Cables | Not needed but included |
Power Source | Rechargable Li ion battery, ext. power supply |
Battery Life | ~ 15 hours |
Battery Tested | According to UN 38.3 |
Exposures Needed for Measurement | One |
Min. Exposure Time | 0.1 ms |
Memory | Unlimited |
Operating Temperature | -15–35°C |
Storage Temperature | -10–50°C |
Waveform Sampling rate | 4–2000 samples/s |
Waveform Recordning time | 1024 ms–524 s |
Backscatter Protection | Yes (both main unit and all detectors) |
Warranty | 2 years |
Extended Warranty | up to 10 years |
Calibration Cycle | 2 years |
Standard | Complies with relevant parts of the standards IEC 61674 and IEC 61676 for Dosimetric instruments. |
EMC Tested | According to IEC 61326-1:2012 |
PTB Approval | DE-17-M-PTB-0070 |
ISO/IEC 17025:2005 Calibration Laboratory
Tube Potential | Mam, R/F, CT, Dental |
Dose | Mam, R/F, CT, Dental |
Dose Area Product | R/F |
Dose Length Product | CT |
Tube Current and Charge | Mam, R/F, CT, Dental |
Photometry | Luminance, Illuminance |
Electrometers | Current, Charge |
Piranha Rad/Flu
Included in main unit | Yes |
Total Filtration | 60–120 kVp, 1–90 mm Al or 2 mm Cu |
Auto Compensation | |
Dose/Dose Rate | 50–150 kVp, 1–90 mm Al or 2 mm Cu |
kVp | 50–150 kVp, 1–90 mm Al or 2 mm Cu |
Dose | |
Range | 0.1 nGy–1500 Gy* |
Inaccuracy | 5% |
Dose Rate | |
Range | 1 nGy/s*–320 mGy/s |
Resolution | 0.2 nGy/s* |
Trig Level | 40 nGy/s |
Inaccuracy | 5% or 0.25 nGy/s |
kVp | |
Range | 35–160 kVp |
Minimum Dose Rate | 0.1 µGy/s |
Inaccuracy | 1.5% |
PPV | Yes |
HVL | |
Range | 0.72–13 mm Al |
Minimum Peak Dose Rate | 0.1 µGy/s |
Inaccuracy | 10% or 0.2 mm Al |
Quick HVL in one exposure | Yes |
Total filtration | |
Range | 1.0–90 mm Al |
Minimum Dose Rate | 0.1 µGy/s |
Inaccuracy | 10% or 0.3 mm Al |
Pulses | |
Range | 1–65535 pulses |
Minimum Dose Rate | 0.23 µGy/s |
Minimum Pulse Width | 0.5 ms at 1.8 µGy/s* |
Pulse Rate | |
Range | 0.5–180 pulse/s |
Minimum Peak Dose Rate | 0.23 µGy/s |
Dose/Pulse | |
Range | 1 nGy/pulse–0.9 Gy/pulse* |
Minimum Peak Dose Rate | 0.23 µGy/s* |
Time | |
Range | 0.1 ms–2000 s |
Resolution | 0.5 ms |
Bandwidth | 1 kHz |
Inaccuracy | 1% or 0.5 ms |
Effective time | Yes (Effective Exposure time) |
Waveforms | |
Resolution | 0.5 ms |
Bandwidth kV | ≤1 kHz |
Bandwidth Dose Rate | ≤1 kHz |
Piranha CT
Weight | 50 g** |
Size | 12.5x12.5x210 mm** |
Size ø | 12.5 mm** |
Energy Dependence | Auto compensated |
Automatic environmental compensation |
Not needed |
Spatial resolution** | 0.3 mm |
Achievable parameters** | CTDI100, CTDIw, CTDIvol, DLP, FWHM, Geometric Efficiency |
CTDI Profile** | Yes** |
Maximum Beam Width | Unlimited** |
Dose | |
Range | 1.8 nGy–22 kGy** |
Inaccuracy | 5%** |
Dose Length Product | |
Range |
1.8 nGycm**–22 kGycm** at 10 mm collimation |
Inaccuracy | 5%** |
Dose Rate | |
Range | 67 µGy/s–2.2 mGy/s** |
Inaccuracy | 5%** |
Time | |
Range | 0.1 ms–3400s |
Resolution | ≥0.5 ms |
Bandwidth | ≤1 kHz |
Inaccuracy | 1% or 0.5 ms |
Waveforms | |
Resolution | ≥0.5 ms |
Bandwidth | ≤1 kHz |
kV | |
Range | 45–155 kVp |
Inaccuracy | 1.5% |
PPV | Yes |
Mammography
Included in main unit | Yes |
Tailormade calibrations for different manufactures |
Yes |
Full mammo range available without adding extra filters or repositioning or changing detector. |
Yes |
Auto Compensation |
|
kV/Dose/Dose Rate & HVL | |
Mo/Mo | 18–49 kVp |
Mo/Rh | 20–46 kVp |
Mo/Al | 18–49 kVp |
Mo/Cu | 40–49 kVp |
Rh/Rh | 25–49 kVp |
Rh/Cu | 40–49 kVp |
W/Rh | 20–49 kVp |
W/Ag | 20–40 kVp |
W/Al | 20–49 kVp |
W/Cu | 40–49 kVp |
Tailormade Calibrations | |
kV/Dose/Dose Rate & HVL | Fujifilm: Amulet innovality |
GE: Pristina | |
IMS: Giotto | |
Hologic: Selenia, Selenia Dimensions, Affirm Prone |
|
Dose | |
Range | 1 nGy–1000 Gy* |
Inaccuracy | 5% |
Dose Rate | |
Range | 25 nGy/s – 530 mGy/s* |
Inaccuracy | 5% |
kVp | |
Range | 18–49 kVp |
Minimum Dose Rate | 0.2 µGy/s |
Inaccuracy | 1.5% or ± 0.7 kV for Mo/Mo |
PPV | Yes |
HVL | |
Range | 0.19–4.3 mmAl |
Minimum Dose Rate | 0.2 µGy/s |
Inaccuracy | 10% |
Quick HVL | Yes |
Time | |
Range | 0.1 ms–2000 s |
Resolution | 0.5 ms |
Bandwidth | 1000 Hz |
Inaccuracy | 1% or 0.5 ms |
Pulses | |
Range | 1–9999 pulses |
Pulse Rate | |
Range | 0.5–180 pulses/s |
Dose/Pulse | |
Range | 8 nGy/pulse–0.9 Gy/pulse* |
Minimum Dose Rate | 0.46 µGy/s |
Waveforms | |
Resolution | 0.5 ms |
Bandwidth | 1000 Hz |
MAS-1
mAs | |
Range | 0.001–9999 mAs |
Resolution | 0.001 mAs |
Inaccuracy | 1% |
mA | |
Range | 0.1–3000 mA |
Resolution | 0.01 mA |
Inaccuracy | 1% or 0.01 mA |
Time | |
Range | 0.1 ms–34000 s |
Resolution | 0.5 ms |
Bandwidth | 500 Hz |
Inaccuracy | 1% or 0.5 ms |
Pulses | |
Range | 1-65535 |
Resolution | 1 pulse |
Inaccuracy | 1 pulse |
Pulse Rate | |
Range | 0.5–100 Hz |
Resolution | 0.5 Hz |
mAs/Pulse | |
Range | 0.001–9999 mAs |
Resolution | 0.001 mAs |
Inaccuracy | 1% |
Waveform | |
Resolution | 0.5 ms |
Bandwidth | 0.5 kHz |
MAS-2
mAs | |
Range | 0.001–9999 mAs |
Resolution | 0.001 mAs |
Inaccuracy | 5% |
mA | |
Range | 10 mA–4000 mA |
Resolution | 1 mA |
Inaccuracy | 5% or 2 mA |
Time | |
Range | 0.1 ms–34000 s |
Resolution | 0.5 ms |
Bandwidth | 1 kHz |
Inaccuracy | 1% or 0.5 ms |
Pulses | |
Range | 1-65535 pulses |
Resolution | 1 pulse |
Inaccuracy | 1 pulse |
Pulse Rate | |
Range | 0.5–100 Hz |
Resolution | 0.5 Hz |
mAs/Pulse | |
Range | 0.1–9999 mAs |
Resolution | 0.01 mAs |
Inaccuracy | 5% |
Waveform | |
Resolution | 0.5 ms |
Bandwidth | 0.5 kHz |
Piranha Survey*
Weight | 85 g |
Size | 45x19x7 mm |
Standard | Complies with IEC standard EN 60601-1-3 |
Active Compensation |
Flat energy response. |
Trig Models | |
Manual | Yes. Manual start / stop of measurement |
Auto | Yes. Trig with threshold 40nGy/s |
Air Kerma | |
Range | 0.1 nGy – 1500 Gy |
Inaccuracy | 5% or 0.25 nGy |
Air Kerma Rate | |
Range | 1 nGy/s–150 mGy/s |
Inaccuracy | 5% or 0.9 µGy/h |
Resolution | 0.2 nGy/s |
Mean Energy | |
Range | 18–160 keV |
Inaccuracy | 5% |
Minimum Dose Rate | 4 mGy/h |
Time | |
Range | 0.1 ms–2000 s |
Resolution | 0.5 ms |
Bandwidth | ≤1 kHz |
Inaccuracy | 1% or 0.5 ms |
Waveform | |
Resolution | 0.5–64 ms |
Bandwidth | ≤1000 Hz |
Minimum Dose Rate | No limitation |
Piranha Light
Weight | 190 g*** |
Size | 79 x 47 x 29 mm*** |
Luminance | |
Range | 0.04–128000 cd/m2 |
Resolution | 0.5 mcd/m2 |
Aperture Angle | 3.6° |
Measurement Area | ø 7 |
Inaccuracy Illuminant | 5% |
Illuminance | |
Range | 0.014–48000 lux |
Resolution | 0.2 mlux |
Uncertainty Illuminant A | 5% |
Deviation from human eye (λ) (f1’) | See figure |
Cosine Deviation (f2) | <3% |
* RTI Dose Probe
** CT Dose Profiler
*** RTI Light Probe
**** RTI T20
Specifications are based on a Piranha model 657.
Article Nr (Piranha 657): 9729657-00
Specifications_Black Piranha.pdf
Black Piranha Selection Guide.pdf
Accessories
RTI Light Probe
Light Detector measures the brightness on monitors and film viewing boxes.
RTI Dose Probe
The RTI Dose Probe is specially designed for low dose rate measurements.
RTI T20 Dose Detector
T20 dose detector is small enough to leave the X-ray system's output undisturbed.
DAP Chamber
The DAP Chamber is the perfect tool for field calibration of orthodontic X-ray equipment. You can measure with just one click – fast, easy and accurate.
Piranha Premium ALU Case
Made of light-weight aluminum, the Piranha Premium ALU Case is attractive and just “the right size” to provide a space for the Piranha and all its accessories included in the popular Piranha Premium Kits.
Piranha Premium Outdoor Case
A tough, dent-resistant, transport for your Piranha Premium System.
RTI Soft Shell Case
RTI's soft shell case is a perfect solution for storage of your Cobia or Piranha. The soft shell case fits perfectly for the Cobia Smart or Piranha and is smooth, easy and safe.
Support
1. Do you have the latest software?
It is always good to have the latest update for your instrument. This is because we constantly fix bugs and make improvements.
2. Have you read our FAQ?
For the most common application and measurement issues, we have collected the answers in one place. This is so that you will get a quick answer to your question.
3. Have you checked the application notes?
In our application notes, you will find answers, solutions and tips for some specific measuring task that may arise. Here you will also find instructions on how to measure on some unusual X-ray machines and new measuring set-up.
4. Contact your local distributor
Your local representative – our extension in the world – is an expert on RTI's products and can help youto answer your questions and support you. This way you can get help in your time zone and in your own language.
5. Contact us
We encourage you to contact us if you have questions or problems. We are here to help you and be your partner were in your everyday working life.
Product Configuration
Which RTI Solution Suits Your Needs? RTI offers complete quality assurance solutions for all types or X-ray equipment.
To be able to find out which instrument and solution that suits you, we need to know a few things about you and your measuring situation. Please fill out this form, and we will get back to you with a product suggestion.
Email sent