    RTI Slit Camera

    Focal Spot Measurements

    The RTI Slit Camera for X-ray focal spot measurements, to assure that the tube and X-ray tube focal spot size is within acceptable limits. The Slit Camera can be rotated 90 degrees in the Slit Camera holder. The 90 degree rotation simplifies the measurement of length and width measurements of the focal spot since the slit only have to be aligned once. 

    RTI Slit CAM

    RTI Slit CAM for Focal Spot

    The RTI Slit Camera for X-ray focal spot measurement comes with a stand with an adjustable rod. Optionally also a tilting device is available. 

    90 Degree Rotation

    The Slit Camera can be rotated 90 degrees in the Slit Camera holder. The 90 degree rotation simplifies the measurement of length and width measurements of the focal spot since the slit only have to be aligned once.

     

    Slit Stand Titling Scale

    Tilting for Mammography

    The optional tilting device enables measurements where the reference position is not orthogonal to the image plane, which is normal in mammography.

    Highlights

     

     

    General  
    Supporting standards

    IEC 60336

    Tube focal spot types

    Radiographic, CT, Mammographic, Dental

    kV range

    18 – 150 kV

    Enlargement factor E

    E = n/m, where:

    n = slit to detector distance

    m = slit to focal spot distance

    0.8 – 5

    Measurement range

    0.1 – 5 mm

    Dimensions

     

    Slit holder size

    97 x 34 x 6 mm

    Slit dimensions

    10 µm (as defined by IEC 60336)

    Slit material

    Tungsten (W, 99.9 %)

       

     

    Article nr: 9707030-00

     