Piranha Multi (657)

$13,200

All you need for qualified diagnostics and QA on radiography, fluoroscopy, CT, dental, and mammography X-ray scans. The Piranha Multi measures kVp, Time, HVL, Total Filtration, Dose, Dose rate, and presents waveform.

Included: RTI Dose Probe (calibrated at 70 kV, W/23 mm Al (R2) – for input dose to image intensifiers, scattered and leakage radiation), 10” Tablet PC, internal rechargeable battery, internal Bluetooth and USB port, USB cable, Ocean software with Connect license, power supply, USB Stick, and soft-shell transport case.

Art. No 9729657-00

All prices include freight cost. Pricing does not include Applicable Taxes or Customs Fees.

The complete solution for QA testing

The Piranha Multi is the complete solution for QA testing in radiography, fluoroscopy, CT, dental, and mammography.

It measures and presents all parameters instantly and simultaneously: kVp; Time; HVL; Total Filtration; Dose; Dose rate; and presents waveforms. Easy & fast X-ray Quality Control!

With the Piranha Multi, you have a compact solution for fast and reliable testing. You can be sure of extremely accurate measurements that can be used for in-depth analysis.

Ocean software – quick and repeatable
Its compatibility with the Ocean diagnostic software provides quick and repeatable QA work. Ocean can perform instant real-time analysis during your measurements. Use your tablet/laptop as both an interactive display during the measurements and as a powerful analysis tool when you are back at the office.

All your measurements, along with the result and the report, will be stored inside your tablet or laptop; there is no unnecessary, time-consuming data transfer.

A complete QA solution instantly
The Piranha Multi comes ready-to-use with everything you need included. It is really an all-in-one multifunction meter. Connect to your computer wireless or via USB and you have a complete QA system!

Position Check function
The unique feature of the Piranha Multi makes it possible to check the position of the detector before measuring. By using the “Position Check” function, you can verify that the detector area is fully irradiated. Possible field inhomogeneities are also neutralized.

Piranha Multi (657) key features

  • All-in-one multifunction X-ray meter
  • One-shot HVL
  • Solid-state detectors = no need to compensate for temperature & pressure
  • Can measure on scanning beams as well as tomosynthesis
  • Optimized for X-ray equipment from a large number of manufacturers
  • Built-in energy compensation
  • Can be used together with ion chambers
  • Wide-range detection of total filtration
  • 100 meters Bluetooth range
  • Automatic recognition of external probes
  • Small, compact, & robust – easy to place
  • Backscatter protected
  • Long-lasting rechargeable battery
  • Always free firmware upgrades
  • Up to 10-year warranty
  • Two-year Calibration Cycle

General specifications Piranha

  • Size
    133x75x26 mm
  • Weight
    Approx. 400 g
  • Display
    PC or Windows Tablet
  • Software
    Ocean - Complete Software for X-ray QA
  • DatabaseSupport
    Yes
  • Interface Type
    Built-in Bluetooth and USB
  • Detector Position Check
    Yes
  • Simultaneous Dose Measurements
    Yes - up to two detectors simultaneously
  • Use of Dose Detector with Active AEC
    Yes - RTI Dose Probe or RTI T20
  • Bluetooth range
    100 m free-in-air
  • USB Cables
    Included
  • Power source
    Rechargable Li ion battery / Ext. power supply
  • Battery life
    Approx. 15 hours
  • Battery tested
    According to UN 38.3
  • No. of Exposures Needed for Measurement
    One
  • Min. Exposure Time
    0.1 ms
  • Memory
    Unlimited
  • Operating temperature
    -15 °c – 35 °c
  • Storage temperature
    -10 °c - +50°c
  • Waveform Sampling rate
    4 – 2000 samples/s
  • Waveform recording time
    1024 ms – 524 s
  • Backscatter Protection
    Yes - main unit, RTI Dose Probe, and T20
  • Warranty
    Two years
  • Extended Warranty
    Up to 10 years
  • Calibration Cycle
    Two years
  • Standard
    Complies with relevant parts of the standards IEC 61674 and IEC 61676 for Dosimetric instruments
  • EMC Tested
    According to IEC 61326-1:2012
  • PTB Approval
    DE-17-M-PTB-0070

Specifications R/F, CT, Dental

  • Auto Compensation
  • Dose/Dose Rate
    50 – 150 kVp, 1 – 90 mm Al, or 2 mm Cu
  • kV
    50–150 kVp, 1–90 mm Al, or 2 mm Cu
  • Total Filtration
    50–150 kVp, 1–90 mm Al, or 2 mm Cu
  • Dose
  • Range
    0.1 nGy – 1500 Gy*
  • Inaccuracy
    5%
  • Dose Rate
  • Range
    4 nGy/s - 320 mGy/s
  • Resolution
    0.2 nGy/s*
  • Trig Level
    40 nGy/s
  • Inaccuracy
    5% or 0.25 nGy/s
  • Tube potential
  • Range
    35–160 kVp
  • Minimum Dose Rate
    0.1 µGy/s
  • Inaccuracy
    1.5%
  • PPV
    Yes
  • HVL
  • Range
    0.72 – 13 mm Al
  • Minimum Peak Dose Rate
    0.1 µGy/s
  • Inaccuracy
    10 % or 0.2 mm Al
  • Quick HVL in one exposure
    Yes
  • Total filtration
  • Range
    1.0–90 mm Al
  • Minimum Dose Rate
    0.1 µGy/s
  • Inaccuracy
    10 % or 0.3 mm Al
  • Pulses
  • Range
    1–65535 pulses
  • Minimum Dose Rate
    0.23 µGy/s
  • Minimum Pulse Width
    0.5 ms at 1.8 µGy/s*
  • Pulse Rate
  • Range
    0.5–180 pulse/s
  • Minimum Peak Dose Rate
    0.23 µGy/s
  • Dose/Pulse
  • Range
    1 nGy/pulse – 0.9 Gy/pulse*
  • Minimum Peak Dose Rate
    0.23 µGy/s*
  • Time
  • Range
    0.1 ms – 2000 s
  • Resolution
    0.5 ms
  • Bandwidth
    1 kHz
  • Inaccuracy
    0.5% or 0.5 ms
  • Effective time
    Yes (Effective Exposure time)
  • Waveforms
  • Resolution
    0.5 ms
  • Bandwidth kV
    ≤1 kHz
  • Bandwidth Dose Rate
    ≤1 kHz

Specifications Mammography

  • Tailor-made calibrations for different manufactures
    Yes
  • Full mammo range available without adding extra filters, repositioning, or changing detector
    Yes

Auto Compensation - kV/Dose/Dose Rate & HVL

  • Mo/Mo
    18 – 49 kVp
  • Mo/Rh
    20 – 46 kVp
  • Mo/Al
    18 – 49 kVp
  • Mo/Cu
    40 – 49 kVp
  • Rh/Rh
    25 – 49 kVp
  • Rh/Ag
    27 - 40 kVp
  • Rh/Al
    22 - 35 kVp
  • Rh/Cu
    40 – 49 kVp
  • W/Rh
    20 – 49 kVp
  • W/Ag
    20 – 40 kVp
  • W/Al
    20 – 49 kVp
  • W/Cu
    40 – 49 kVp
  • W/Ti
    40 – 49 kVp
  • Tailor-made Calibrations
  • kV/Dose/Dose Rate, & HVL
    Fujifilm, Amulet Innovality/Cristal
    GE, Pristina
    IMS Giotto
    Hologic Selenia, Selenial Dimensions, Affirm Prone
    Siemens Revelation
  • Dose
  • Range
    1 nGy – 1000 Gy*
  • Inaccuracy
    5%
  • Dose Rate
  • Range
    25 nGy/s – 530 mGy/s*
  • Inaccuracy
    5%
  • kVp
  • Range
    18–49 kVp
  • Minimum Dose Rate
    0.2 µGy/s
  • Inaccuracy
    1.5% or ± 0.7 kV for W / 0.7 mm Al
  • PPV
    Yes
  • HVL
  • Range
    0.19–4.3 mmAl
  • Minimum Dose Rate
    0.2 µGy/s
  • Inaccuracy
    10%
  • Quick HVL
    Yes
  • Time
  • Range
    0.1 ms–2000 s
  • Resolution
    0.5 ms
  • Bandwidth
    1 kHz
  • Inaccuracy
    0.5% or 0.5 ms
  • Pulses
  • Range
    1 – 9999 pulses
  • Pulse Rate
  • Range
    0.5 – 180 pulses/s
  • Dose/Pulse
  • Range
    8 nGy/pulse – 0.9 Gy/pulse*
  • Minimum Dose Rate
    0.46 µGy/s
  • Waveforms
  • Resolution
    0.5 ms
  • Bandwidth
    1 kHz

ISO/IEC 17025:2005 Calibration Laboratory

  • Tube Potential
    Mam, R/F, CT, Dental
  • Dose
    Mam, R/F, CT, Dental
  • Dose Area Product
    R/F
  • Dose Length Product
    CT
  • Tube Current and Charge
    Mam, R/F, CT, Dental
  • Photometry
    Luminance, Illuminance
  • Electrometers
    Current, Charge

Notice

  • * With RTI Dose Probe

Probes

Holders/Stands