Piranha R/F (557)

$10,800

A compact solution for fast and reliable testing covering all your QA requirements. Presents all parameters instantly and simultaneously for Radiography, Fluoroscopy, CT, and Dental. Measures kVp, Time, HVL, Total Filtration, Dose, Dose rate, and presents waveform.

Included: RTI Dose Probe (calibrated at 70 kV, W/23 mm Al – for input dose to image intensifiers, scattered and leakage radiation), 10” Tablet PC, internal rechargeable battery, internal Bluetooth and USB port, USB cable, Ocean software with Connect license, power supply, USB Stick, and soft-shell transport case.

Art. No 9729557-00

All prices include freight cost. Pricing does not include Applicable Taxes or Customs Fees.

Fast and reliable testing

With the Piranha R/F (557), you’ll have all you need for advanced quality control – just “Plug n Play”.

Piranha is easy to use with time-saving features. All the necessary measurement parameters are covered; kVp, Time, HVL, Total Filtration, Dose, Dose rate, and presents waveforms.

Measurements are transmitted to your Ocean software for analysis, storage, and compiling reports in one smooth process. Fast and reliable, Piranha gives you extremely accurate measurements for all your in-depth analysis.

With the Piranha R/F meter, you have a compact solution for fast and reliable testing. You can be sure of extremely accurate measurements that can be used for in-depth analysis.

Ocean software – quick and repeatable
Its compatibility with the Ocean diagnostic software provides quick and repeatable QA work. Ocean can perform instant real-time analysis during your measurements. Use your tablet/laptop as both an interactive display during the measurements and as a powerful analysis tool when you are back at the office.

All your measurements, along with the result and the report, will be stored inside your tablet or laptop; there is no unnecessary, time-consuming data transfer.

A complete QA solution instantly
The Piranha R/F comes ready-to-use with everything you need included. It is really an all-in-one multifunction meter. Connect to your computer wireless or via USB and you have a complete QA system!

Position Check function
The unique feature of the Piranha R/F makes it possible to check the position of the detector before measuring. By using the “Position Check” function, you can verify that the detector area is fully irradiated. Possible field inhomogeneities are also neutralized.

Piranha R/F (557) key features

  • All-in-one multifunction X-ray meter
  • One-shot HVL
  • Solid-state detectors = no need to compensate for temperature & pressure
  • Optimized for X-ray equipment from a large number of manufacturers
  • Built-in energy compensation
  • Can be used together with ion chambers
  • Wide-range detection of total filtration
  • 100 meters Bluetooth range
  • Unique detector design to minimize position and rotation dependence
  • Automatic recognition of external probes
  • Small, compact, & robust – easy to place
  • Backscatter protected
  • Long-lasting rechargeable battery
  • Always free firmware upgrades
  • Upgradable to any Piranha model
  • Up to 10-year warranty
  • Two-year Calibration Cycle

General specifications Piranha

  • Size
    133x75x26 mm
  • Weight
    Approx. 400 g
  • Display
    PC or Windows Tablet
  • Software
    Ocean - Complete Software for X-ray QA
  • DatabaseSupport
    Yes
  • Interface Type
    Built-in Bluetooth and USB
  • Detector Position Check
    Yes
  • Simultaneous Dose Measurements
    Yes - up to two detectors simultaneously
  • Use of Dose Detector with Active AEC
    Yes - RTI Dose Probe or RTI T20
  • Bluetooth range
    100 m free-in-air
  • USB Cables
    Included
  • Power source
    Rechargable Li ion battery / Ext. power supply
  • Battery life
    Approx. 15 hours
  • Battery tested
    According to UN 38.3
  • No. of Exposures Needed for Measurement
    One
  • Min. Exposure Time
    0.1 ms
  • Memory
    Unlimited
  • Operating temperature
    15 °c – 35 °c
  • Storage temperature
    -10 °c - +50°c
  • Waveform Sampling rate
    4 – 2000 samples/s
  • Waveform recording time
    1024 ms – 524 s
  • Backscatter Protection
    Yes - main unit, RTI Dose Probe, and T20
  • Warranty
    Two years
  • Extended Warranty
    Up to 10 years
  • Calibration Cycle
    Two years
  • Standard
    Complies with relevant parts of the standards IEC 61674 and IEC 61676 for Dosimetric instruments
  • EMC Tested
    According to IEC 61326-1:2012
  • PTB Approval
    DE-17-M-PTB-0070

Specifications R/F, CT, Dental

  • Auto Compensation
  • Dose/Dose Rate
    50 – 150 kVp, 1 – 90 mm Al, or 2 mm Cu
  • kV
    50–150 kVp, 1–90 mm Al, or 2 mm Cu
  • Total Filtration
    50–150 kVp, 1–90 mm Al, or 2 mm Cu
  • Dose
  • Range
    0.1 nGy – 1500 Gy*
  • Inaccuracy
    5%
  • Dose Rate
  • Range
    1 nGy/s* – 320 mGy/s
  • Resolution
    0.2 nGy/s*
  • Trig Level
    40 nGy/s
  • Inaccuracy
    5% or 0.25 nGy/s
  • Tube potential
  • Range
    35–160 kVp
  • Minimum Dose Rate
    0.1 µGy/s
  • Inaccuracy
    1.5%
  • PPV
    Yes
  • HVL
  • Range
    0.72 – 13 mm Al
  • Minimum Peak Dose Rate
    0.1 µGy/s
  • Inaccuracy
    10% or 0.2 mm Al
  • Quick HVL in one exposure
    Yes
  • Total filtration
  • Range
    1.0 – 90 mm Al
  • Minimum Dose Rate
    0.1 µGy/s
  • Inaccuracy
    10% or 0.3 mm Al
  • Pulses
  • Range
    1 – 65535 pulses
  • Minimum Dose Rate
    0.23 µGy/s
  • Minimum Pulse Width
    0.5 ms at 1.8 µGy/s*
  • Pulse Rate
  • Range
    0.5 – 180 pulse/s
  • Minimum Peak Dose Rate
    0.23 µGy/s
  • Dose/Pulse
  • Range
    1 nGy/pulse – 0.9 Gy/pulse*
  • Minimum Peak Dose Rate
    0.23 µGy/s*
  • Time
  • Range
    0.1 ms – 2000 s
  • Resolution
    0.5 ms
  • Bandwidth
    1 kHz
  • Inaccuracy
    0.5% or 0.5 ms
  • Effective time
    Yes (Effective Exposure time)
  • Waveforms
  • Resolution
    0.5 ms
  • Bandwidth kV
    ≤1 kHz
  • Bandwidth Dose Rate
    ≤1 kHz

ISO/IEC 17025:2005 Calibration Laboratory

  • Tube Potential
    Mam, R/F, CT, Dental
  • Dose
    Mam, R/F, CT, Dental
  • Dose Area Product
    R/F
  • Dose Length Product
    CT
  • Tube Current and Charge
    Mam, R/F, CT, Dental
  • Photometry
    Luminance, Illuminance
  • Electrometers
    Current, Charge

Notice

  • * With RTI Dose Probe

Probes

Holders/Stands