Piranha Premium
$19,055.00
All you need for qualified diagnostics and QA on Radiography, Fluoroscopy, CT, Dental, and Mammography X-ray scans. The Piranha Multi measures kVp, Time, HVL, Total Filtration, Dose, Dose rate, and presents waveform.
The Premium Kit is based on a Piranha MULTI along with a number of its probes and accessories, well-organized and safely stored in a robust Aluminum Case.
Read about our pre-paid calibration and extended warranty.
All prices include freight cost. Pricing does not include Applicable Taxes or Customs Fees.
The complete solution for QA testing
The Piranha MULTI is the complete solution for QA testing in radiography, fluoroscopy, CT, dental, and mammography.
It measures and presents all parameters instantly and simultaneously: kVp; Time; HVL; Total Filtration; Dose; Dose rate; and presents waveforms. Easy & fast X-ray Quality Control!
With the Piranha Multi, you have a compact solution for fast and reliable testing. You can be sure of extremely accurate measurements that can be used for in-depth analysis.
Ocean Next™ software
Its compatibility with our Ocean Next™ software provides quick and repeatable QA work. Ocean can perform instant real-time analysis during your measurements. Use your tablet/laptop as both an interactive display during the measurements and as a powerful analysis tool when you are back at the office.
All your measurements, along with the result and the report, will be stored inside your tablet or laptop; there is no unnecessary, time-consuming data transfer.
A complete QA solution instantly
The Piranha Multi comes ready-to-use with everything you need included. It is really an all-in-one multifunction meter. Connect to your computer wireless or via USB and you have a complete QA system!
Position Check function
The unique feature of the Piranha MULTI makes it possible to check the position of the detector before measuring. By using the “Position Check” function, you can verify that the detector area is fully irradiated. Possible field inhomogeneities are also neutralized.
Piranha MULTI key features
- All-in-one multifunction X-ray meter
- One-shot HVL
- Solid-state detectors = no need to compensate for temperature & pressure
- Can measure on scanning beams as well as tomosynthesis
- Optimized for X-ray equipment from a large number of manufacturers
- Built-in energy compensation
- Can be used together with ion chambers
- Wide-range detection of total filtration
- 100 meters Bluetooth range
- Unique detector design to minimize position and rotation dependence
- Automatic recognition of external probes
- Small, compact, and robust – easy to place
- Backscatter protected
- Long-lasting rechargeable battery
- Always free firmware upgrades
- Up to ten-year warranty
- Two-year Calibration Cycle
General specifications Piranha
-
Size133x75x26 mm
-
WeightApprox. 400 g
-
DisplayPC or Windows Tablet
-
SoftwareOcean - Complete Software for X-ray QA
-
DatabaseSupportYes
-
Interface TypeBuilt-in Bluetooth and USB
-
Detector Position CheckYes
-
Simultaneous Dose MeasurementsYes - up to two detectors simultaneously
-
Use of Dose Detector with Active AECYes - RTI Dose Probe or RTI T20
-
Bluetooth range100 m free-in-air
-
USB CablesIncluded
-
Power sourceRechargable Li ion battery / Ext. power supply
-
Battery lifeApprox. 15 hours
-
Battery testedAccording to UN 38.3
-
No. of Exposures Needed for MeasurementOne
-
Min. Exposure Time0.1 ms
-
MemoryUnlimited
-
Operating temperature15 °c – 35 °C
-
Storage temperature-10 °C - +50°C
-
Waveform Sampling rate4 – 2000 samples/s
-
Waveform recording time1024 ms – 524 s
-
Backscatter ProtectionYes - main unit, RTI Dose Probe, and T20
-
WarrantyTwo years
-
Extended WarrantyUp to 10 years
-
Calibration CycleTwo years
-
StandardComplies with relevant parts of the standards IEC 61674 and IEC 61676 for Dosimetric instruments
-
EMC TestedAccording to IEC 61326-1:2012
-
PTB ApprovalDE-17-M-PTB-0070
Piranha - Specifications for Radiography & Fluoroscopy
-
Tube Voltage35 – 160 kV (±1.5 %)
-
PPVYes
-
Time0.1ms – 2000 s (±0.5 % or ±0.5 ms)
-
Dose1.3 nGy – 650 Gy. 150 nR – 74 kR. (±5 %)
-
Dose Rate15 nGy/s – 320 mGy/s (±5 % or ±7 nGy/s)
1.7 μR/s – 37 R/s (±5 % or ±0.8 μR/s) -
Auto CompensationAll dose parameters are automatically compensated for using measured kVp and TF over their specified ranges.
-
HVL1.2 – 14 mm Al (±10 % or ±0.2 mm)5%
-
Total Filtration1.0 – 90 mm Al (±10 % or ±0.3 mm)
-
Pulses1 – 65535 pulses (±1 pulse)
-
Dose/pulse8 nGy/pulse – 0.9 Gy/pulse
-
Pulse Rate/Frequency0.5 – 180 Hz
-
Pulse width4 ms – 2000 s
Piranha - Specifications for Mammography
-
Tube Voltage
-
Target/Filter: Mo / 30 μm Mo
Range: 18 – 49 kV
Inaccuracy: ±0.5 kV or ±1.5 % -
Target/Filter: Mo / 25 μm Rh
Range: 20 – 46 kV
Inaccuracy: ±1 kV -
Target Filter: Rh / 25 μm Rh
Range: 25 – 49 kV
Inaccuracy: ±1 kV -
Target Filter: Rh / 1.0 mm Al
Range: 22 – 35 kV
Inaccuracy: ±1 kV -
Target Filter: W / 50 μm Rh
Range: 20 – 49 kV
Inaccuracy: ±0.5 kV or ±1.5 %0, -
Target Filter: W / 0.50 mm Al
Range: 20 – 48 kV
Inaccuracy: ±1 kV -
Target Filter: Mo / 1.0 mm AI
Range: 18 – 49 kV
Inaccuracy: ±1 kV -
Target Filter: W / 50 μm Ag
Range: 20 – 40 kV
Inaccuracy: ±0.5 kV or ±1.5 % -
Target Filter: W / 75 μm Ag
Range: 20 – 40 kV
Inaccuracy: ±1 kV -
Target Filter: W / 50 μm Rh (Gio)
Range: 22 – 35 kV
Inaccuracy: ±1 kV -
Target Filter: W / 50 μm Rh (Gio)
Range: 22 – 35 kV
Inaccuracy: ±1 kV -
Target Filter: W / 0.70 mm Al
Range: 20 – 49 kV
Inaccuracy: ±0.5 kV or ±1.5 % -
Target Filter: W / 50 μ m Ag (Sel)
Range: 22 – 39 kV
Inaccuracy: ±0.5 kV or ±1.5 % -
Target Filter: W / 50 μ m Rh (Sel)
Range: 22 – 39 kV
Inaccuracy: ±0.5 kV or ±1.5 % -
Target Filter: W / 0.30 mm Cu
Range: 40 – 49 kV
Inaccuracy: ±0.5 kV or ±1.5 % -
Target Filter: W / 0.70 mm Al (Inno/Crist)
Range: 20 – 49 kV
Inaccuracy: ±0.5 kV or ±1.5 % -
Target Filter: W / 50 μ m Rh (Inno/Crist)
Range: 20 – 49 kV
Inaccuracy: ±0.5 kV or ±1.5 % -
Target Filter: Mo / 25 μ m Rh (Sel)
Range: 20 – 46 kV
Inaccuracy: ±0.5 kV or ±1.5 % -
Target Filter: Rh / 30 μm Ag (GE)
Range: 27 – 40 kV
Inaccuracy: ±0.5 kV or ±1.5 % -
Target Filter: Rh / Ag IQST (GE)
Range: 31 – 37 kV
Inaccuracy: ±0.5 kV or ±1.5 % -
Target Filter: Mo / 0.25 mm Cu (GE)
Range: 40 – 49 kV
Inaccuracy: ±0.5 kV or ±1.5 % -
Target Filter: Rh / 0.25 mm Cu (GE)
Range: 40 – 49 kV
Inaccuracy: ±0.5 kV or ±1.5 % -
Target Filter:Mo / 30 μm Mo (GE)
Range: 22 – 32 kV
Inaccuracy: ±1 kV -
Target Filter: Affirm Prone W / 50 μm Ag
Range: 20 – 40 kV
Inaccuracy: ±0.5 kV or ±1.5 % -
Target Filter: Affirm Prone W / 0.70 mm Al
Range: 20 – 49 kV
Inaccuracy: ±0.5 kV or ±1.5 % -
Time0.1 ms – 2000 s (±0.5 % or ±0.5 ms)
-
Dose1 nGy – 1000 Gy. 3 µR – 150 kR (±5 %)
-
Dose Rate25 nGy/s – 530 mGy/s (±5 % or ±12 nGy/s)
30 μR/s – 60 R/s (±5 % or ±1.5 μR/s) -
HVL Range0.19 – 4.3 mm Al (depending on Target/Filter combination) (±10 %)
-
Pulses1 – 65535 pulses (±1 pulse)
Piranha - Specifications for CT
-
Tube Voltage45 – 155 kV (±1.5 %)
-
Time0.1 ms – 2000 s (±0.5 % or ±0.5 ms)
-
Total Filtration1 – 90 mm Al (±10 % or ±0.3 mm)
-
HVL1.2 – 14 mm Al (±10 % or ±0.2 mm)
-
Pulses1 – 65535 pulses (±1 pulse)
Piranha - Specifications for Dental
-
Tube Voltage35 – 125 kV (±1.5 %)
-
PPVYes
-
Time0.1ms – 2000 s (±0.5 % or ±0.5 ms)
-
Dose1.3 nGy – 650 Gy. 150 nR – 74 kR. (±5 %)
-
Dose Rate15 nGy/s – 320 mGy/s (±5 % or ±7 nGy/s). 1.7 μR/s – 37 R/s (±5 % or ±0.8 μR/s)
-
Auto CompensationAll dose parameters are automatically compensated for using measured kVp and TF over their specified ranges.
-
HVL1.2 – 14 mm Al (±10 % or ±0.2 mm)
-
Total Filtration1.0 – 90 mm Al (±10 % or ±0.3 mm)
-
Pulses1 – 65535 pulses (±1 pulse)
-
Dose/pulse8 nGy/pulse – 0.9 Gy/pulse
-
Pulse Rate/Frequency0.5 – 180 Hz
-
Pulse width4 ms – 2000 s
Piranha - Specifications with RTI Dose Probe
-
Size20 x 45 x 7.4 mm, 0.79" x 1.8" x 0.29"
-
Weight85 g (3 oz)
-
Cable length2.0 m (6.6 ft)
-
Backscatter protectedYes
-
Dose100 pGy – 1.5 kGy. 12 nR – 170 kR (± 5%)
-
Dose Rate40 nGy/s – 150 mGy/s (±5 % or ±10 nGy/s)
4.6 μR/s – 16 R/s (±5 % or ±1 μR/s)
260 μR/min – 1000 R/min (±5 % or ±6 μR/min) -
Time0.1 ms – 34000 s (±1 % or ±0.5 ms)
-
Pulses1 – 65535 pulses (±1 pulse)
-
Dose per pulse1 nGy/pulse – 0.3 Gy/pulse
-
Pulse rate/frequency0.5 – 100 Hz
-
Pulse width4 ms - 2000 s
Piranha - Specifications with MAS-1
-
TypeMAS-1B
-
Current range0.1–3000 mA
-
Inaccuracy±1 % or ±0.01 mA
-
Frequency rangeDC–500 Hz (-3 dB)
-
Size60 x 35 x 72 mm
-
Weight125 g
-
Battery lifearound 100 h
-
Rechargeable batterywith mini-USB connector
-
Cable length4 m (extension available)
-
ConnectorHirose
Piranha - Specifications with MAS-2
-
TypeMAS-2
-
Current range10 - 4000 mA
-
Inaccuracy±5 % or ±2 mA (±3 % at 250 mA)
-
Charge range0.1 mAs - 9999 mAs
-
Size183 x 61 x 36 mm
-
ConnectorHirose
-
Maximum cable diameter24 mm
-
Power source9 V "6LR61" Alkaline battery
Piranha - Specifications with CT Dose Profiler
-
Dose rate range67 nGy/s-2.2 Gy/s. 0.46 mR/min - 15000 R/min
-
Inaccuracy±5 % or ±15 nGy/s
-
Spatial resolution0.25 mm
Piranha - Specifications with CT Ion Chamber
-
Air kerma rate0.3 mGycm/s to 3 Gycm/s
-
Inaccuracy±5 % or ±0.03 mGycm/s
-
Connector typeLEMO triaxial
-
Cable2 m, low noise triaxial
-
Active volume16 cm³
-
Active length300 mm
-
Diameter12 mm, 12.6 incl O-rings
-
Typical leakage±20 fA
-
Radiation quality70 - 150 kV
-
Sensitivity30 mGycm/nC
-
Energy Dependence±1 %