Fluoroscopy Phantom FLUORO-X-4-150
Fluoroscopy image quality phantom
The FLUORO-X-4-150 Phantom Kit is used to check image quality performance of digital and conventional fluoroscopy X-ray systems using single exposure.
These image quality tests are performed on the acceptance test of an X-ray machine to establish a baseline, and on a routine basis. By keeping an on-going record of the test results, it is easy to find any deterioration in performance.
FLUORO-X-4-150 should be used with Cu or Al attenuator plates, included in the kit. The following image quality tests can be performed in a robust repeatable fashion:
- Patient dose – requires additional dosemeter
- Dose indicator
- Limiting spatial resolution
- Contrast resolution
- Effective radiation field
- Light field to X-ray field alignment
- Beam Centering
- Artifacts
Test equipment included in the set:
Fluoro-X-4-150 phantom
Beam centering tower
25 mm Al attenuator (Patient-equivalent)
1.0 mm Cu attenuator
0.5 mm Cu attenuator
Vertical hanger-type holder
Fluoroscopy stand
USB stick with product manuals
Protective case
Designed in accordance with standards DIN 6868-4 & DIN 6868-150
Download the Specification sheet
$3,990