RTI Dose Probe

For low dose rate measurements

 

The RTI Dose Probe is an external dose probe designed to perform very low dose rate measurements, for instance, on image intensifiers.

The probe is very small to avoid or minimize interference with AEC (Automatic Exposure Control) on X-ray equipment. This also enables it to fit into the table bucky.

Since it is a solid-state detector, no corrections for pressure or temperature are needed. Neither is polarizing voltage.

 

Dose probe for Fluoroscopy

The RTI Dose Probe has a fast response which makes it ideal for pulsed fluoroscopy. It can detect individual pulses, determine pulse rate, and show waveform at dose rate.

 

All prices include freight cost. Pricing does not include Applicable Taxes or Customs Fees. Add Voucher Code in the cart.

$1,600

  1. Size 20 x 45 x 7.4 mm, 0.79" x 1.8" x 0.29"
    Weight 85 g (3 oz)
    Cable length 2.0 m (6.6 ft)
    Backscatter protected Yes
    Dose 100 pGy – 1.5 kGy. 12 nR – 170 kR (± 5%)
    Dose rate 40 nGy/s – 150 mGy/s (±5 % or ±10 nGy/s).
    4.6 μR/s – 16 R/s (±5 % or ±1 μR/s).
    260 μR/min – 1000 R/min (±5 % or ±6 μR/min).
    Time 0.1 ms – 34000 s (±1 % or ±0.5 ms)
    Pulses 1 – 65535 pulses (±1 pulse)
    Dose per pulse 1 nGy/pulse – 0.3 Gy/pulse
    Pulse rate/frequency 0.5 – 100 Hz
    Pulse width 4 ms - 2000 s
  2. Size 20 x 45 x 7.4 mm, 0.79" x 1.8" x 0.29"
    Weight 85 g (3 oz)
    Cable length 2.0 m (6.6 ft)
    Backscatter protected Yes
    Dose 6 nGy - 2 kGy. 660 nR - 250 kR (± 5%)
    Dose rate 220 nGy/s - 220 mGy/s. 1.5 mR/min - 1.5 kR/min (± 5%)
    Time 0.33 ms – 9999 s
    Pulses 3 – 9999 pulses
    Dose per pulse 1 nGy/pulse - 3 kGy/pulse. 114 nR/pulse - 342 kR/pulse
    Pulse rate/frequency 1/6 - 260 Hz