Piranha MULTI

The complete solution for X-ray QA Testing

The Piranha MULTI X-ray meter is all you need for qualified diagnostics and QA on Rad/Fluoro, CT, Dental, and Mammography X-ray scans.

Measures kVp, Time, HVL, Total Filtration, Dose, Dose rate, presents Waveform, and much more. See specifications for further information. Customize your measurements and reports with Ocean Next™ software.

Included: RTI Dose Probe (calibrated at 70 kV, W/23 mm Al (R2) – for input dose to image intensifiers, scattered and leakage radiation), internal rechargeable battery, internal Bluetooth and USB port, USB cable, Ocean Next™ software with Advantage license, power supply, USB Stick, Soft-Shell transport case, and Basic online training.

Standard two-year warranty/calibration. Read about our pre-paid calibration and extended warranty.

All prices include freight cost.

Prices do not include applicable taxes or customs fees. Please check your local import requirements.

Add Voucher Code in the cart.

$17,325

A complete QA solution

The RTI Piranha MULTI X-ray meter comes ready to use with everything you need. It is really an all-in-one multifunction meter. Connect to your computer wireless or via USB for a complete QA system.

Position Check function

The unique feature of the RTI Piranha MULTI X-ray meter makes it possible to check the position of the detector before measuring. By using the Position Check function, you can verify that the detector area is fully irradiated. Possible field inhomogeneities are also neutralized.

Ocean Next™ software

Its compatibility with our Ocean Next™ software provides quick and repeatable QA work. Ocean can perform instant real-time analysis during your measurements. Use your tablet/laptop as both an interactive display during the measurements and as a powerful analysis tool when you are back at the office

  1. All-in-one multifunction X-ray meter
  2. One-shot HVL
  3. Solid-state detectors = no need to compensate for temperature & pressure
  4. Can measure on scanning beams as well as tomosynthesis
  5. Optimized for X-ray equipment from a large number of manufacturers
  6. Built-in energy compensation
  7. Can be used together with ion chambers
  8. Wide-range detection of total filtration
  9. 100 meters Bluetooth range
  10. Automatic recognition of external probes
  11. Small, compact, & robust – easy to position
  12. Backscatter protected
  13. Long-lasting rechargeable battery
  14. Always free firmware upgrades
  15. Up to ten-year warranty
  16. Two-year Calibration Cycle
  1. Size 133 x 75 x 26 mm
    Weight Approx. 400 g
    Display PC or Windows Tablet
    Software Ocean Next™ software for X-ray QA
    Database support Yes
    Interface type Built-in Bluetooth and USB
    Detector position check Yes
    Simultaneous dose measurements Yes - up to two detectors simultaneously
    Use of Dose Detector with Active AEC Yes - RTI Dose Probe or RTI T20
    Bluetooth range 100 m free-in-air
    USB cables Included
    Power source Rechargeable Li ion battery / Ext. power supply
    Battery life Approx. 15 hours
    Battery tested According to UN 38.3
    No. of exposures needed for measurement One
    Min. exposure time 0.1 ms
    Memory Unlimited
    Operating temperature 15 °C – 35 °C
    Storage temperature -10 °c - +50°c
    Waveform sampling rate 4 – 2000 samples/s
    Waveform recording time 1024 ms – 524 s
    Backscatter protection Yes - main unit, RTI Dose Probe, and RTI T20
    Warranty Two years
    Extended Warranty Up to ten years
    Calibration cycle Two years
    Standard Complies with relevant parts of the standards IEC 61674 and IEC 61676 for Dosimetric instruments
    EMC Tested According to IEC 61326-1:2012
    PTB Approval DE-17-M-PTB-0070
  2. Tube Voltage 35 – 160 kV (±1.5 %)
    PPV Yes
    Time 0.1ms – 2000 s (±0.5 % or ±0.5 ms)
    Dose 1.3 nGy – 1500 Gy. 150 nR – 150 kR. (±5 %)
    Dose Rate 15 nGy/s – 320 mGy/s (±5 % or ±7 nGy/s)
    1.7 μR/s – 37 R/s (±5 % or ±0.8 μR/s)
    Auto Compensation All dose parameters are automatically compensated for using measured kVp and TF over their specified ranges.
    HVL 0.72 – 13 mm Al (±10 % or ±0.2 mm)
    Total Filtration 1.0 – 90 mm Al (±10 % or ±0.3 mm)
    Pulses 1 – 65535 pulses (±1 pulse)
    Dose/Pulse 8 nGy/pulse – 0.9 Gy/pulse
    Pulse rate/frequency 0.5 – 180 Hz
    Pulse width 4 ms – 2000 s
  3. Tube Voltage
    Target Filter: W / Ti
    Range: 40-49kV
    Inaccuracy: ±0.5kV or ±1.5%
    Target/Filter: Mo / 30 μm Mo
    Range: 18 – 49 kV
    Inaccuracy: ±0.5 kV or ±1.5 %
    Target/Filter: Mo / 25 μm Rh
    Range: 20 – 46 kV
    Inaccuracy: ±1 kV
    Target Filter: Rh / 25 μm Rh
    Range: 25 – 49 kV
    Inaccuracy: ±1 kV
    Target Filter: Rh / 1.0 mm Al
    Range: 22 – 35 kV
    Inaccuracy: ±1 kV
    Target Filter: W / 50 μm Rh
    Range: 20 – 49 kV
    Inaccuracy: ±0.5 kV or ±1.5 %
    Target Filter: W / 0.50 mm Al
    Range: 20 – 48 kV
    Inaccuracy: ±1 kV
    Target Filter: Mo / 1.0 mm AI
    Range: 18 – 49 kV
    Inaccuracy: ±1 kV
    Target Filter: W / 50 μm Ag
    Range: 20 – 40 kV
    Inaccuracy: ±0.5 kV or ±1.5 %
    Target Filter: W / 75 μm Ag
    Range: 20 – 40 kV
    Inaccuracy: ±1 kV
    Target Filter: W / 50 μm Rh (Gio)
    Range: 22 – 35 kV
    Inaccuracy: ±1 kV
    Target Filter: W / 0.70 mm Al
    Range: 20 – 49 kV
    Inaccuracy: ±0.5 kV or ±1.5 %
    Target Filter: W / 50 μ m Ag (Sel)
    Range: 22 – 39 kV
    Inaccuracy: ±0.5 kV or ±1.5 %
    Target Filter: W / 50 μ m Rh (Sel)
    Range: 22 – 39 kV
    Inaccuracy: ±0.5 kV or ±1.5 %
    Target Filter: W / 0.30 mm Cu
    Range: 40 – 49 kV
    Inaccuracy: ±0.5 kV or ±1.5 %
    Target Filter: W / 0.70 mm Al (Inno/Crist)
    Range: 20 – 49 kV
    Inaccuracy: ±0.5 kV or ±1.5 %
    Target Filter: W / 50 μ m Rh (Inno/Crist)
    Range: 20 – 49 kV
    Inaccuracy: ±0.5 kV or ±1.5 %
    Target Filter: Mo / 25 μ m Rh (Sel)
    Range: 20 – 46 kV
    Inaccuracy: ±0.5 kV or ±1.5 %
    Target Filter: Rh / 30 μm Ag (GE)
    Range: 27 – 40 kV
    Inaccuracy: ±0.5 kV or ±1.5 %
    Target Filter: Rh / Ag IQST (GE)
    Range: 31 – 37 kV
    Inaccuracy: ±0.5 kV or ±1.5 %
    Target Filter: Mo / 0.25 mm Cu (GE)
    Range: 40 – 49 kV
    Inaccuracy: ±0.5 kV or ±1.5 %
    Target Filter: Rh / 0.25 mm Cu (GE)
    Range: 40 – 49 kV
    Inaccuracy: ±0.5 kV or ±1.5 %
    Target Filter:Mo / 30 μm Mo (GE)
    Range: 22 – 32 kV
    Inaccuracy: ±1 kV
    Target Filter: Affirm Prone W / 50 μm Ag
    Range: 20 – 40 kV
    Inaccuracy: ±0.5 kV or ±1.5 %
    Target Filter: Affirm Prone W / 0.70 mm Al
    Range: 20 – 49 kV
    Inaccuracy: ±0.5 kV or ±1.5 %
    Time 0.1 ms – 2000 s (±0.5 % or ±0.5 ms)
    Dose 1 nGy – 1000 Gy. 3 µR – 150 kR (±5 %)
    Dose Rate 25 nGy/s – 530 mGy/s (±5 % or ±12 nGy/s)
    30 μR/s – 60 R/s (±5 % or ±1.5 μR/s)
    HVL Range 0.19 – 4.3 mm Al (depending on Target/Filter combination) (±10 %)
    Pulses 1 – 65535 pulses (±1 pulse)
  4. Tube Voltage 45 – 155 kV (±1.5 %)
    Time 0.1 ms – 2000 s (±0.5 % or ±0.5 ms)
    Total Filtration 1 – 90 mm Al (±10 % or ±0.3 mm)
    HVL 0.72 – 13 mm Al (±10 % or ±0.2 mm)
    Pulses 1 – 65535 pulses (±1 pulse)
  5. Tube Voltage 35 – 125 kV (±1.5 %)
    PPV Yes
    Time 0.1ms – 2000 s (±0.5 % or ±0.5 ms)
    Dose 1.3 nGy – 650 Gy. 150 nR – 74 kR. (±5 %)
    Dose Rate 15 nGy/s – 320 mGy/s (±5 % or ±7 nGy/s)
    1.7 μR/s – 37 R/s (±5 % or ±0.8 μR/s)
    Auto Compensation All dose parameters are automatically compensated for using measured kVp and TF over their specified ranges.
    HVL 0.72 – 13 mm Al (±10 % or ±0.2 mm)
    Total Filtration 1.0 – 90 mm Al (±10 % or ±0.3 mm)
    Pulses 1 – 65535 pulses (±1 pulse)
    Dose/pulse 8 nGy/pulse – 0.9 Gy/pulse
    Pulse rate/frequency 0.5 – 180 Hz
    Pulse width 4 ms – 2000 s
  6. Size 20 x 45 x 7.4 mm, 0.79" x 1.8" x 0.29"
    Weight 85 g (3 oz)
    Cable length 2.0 m (6.6 ft)
    Backscatter protected Yes
    Dose 100 pGy – 1.5 kGy. 12 nR – 170 kR (± 5%)
    Dose rate 4 nGy/s – 150 mGy/s (±5 % or ±10 nGy/s).
    0.46 μR/s – 16 R/s (±5 % or ±1 μR/s).
    26 μR/min – 1000 R/min (±5 % or ±6 μR/min).
    Time 0.1 ms – 34000 s (±1 % or ±0.5 ms)
    Pulses 1 – 65535 pulses (±1 pulse)
    Dose per pulse 1 nGy/pulse – 0.3 Gy/pulse
    Pulse rate/frequency 0.5 – 100 Hz
    Pulse width 4 ms - 2000 s

Piranha Premium ALU Case

A solid aluminum case The lightweight aluminum Piranha Premium ALU Case is not only attractive, but it is also just “the right size” to provide space for the Piranha and all its accessories included in the popular Piranha Premium Kits. It gives the Piranha user the perfect solution for transporting and storing the RTI X-ray [...]

Read more

LoniMover™

Measure CTDI for Wide Beam and Cone Beam CT! Together with the RTI CT Ion Chamber or CT Dose Profiler, the RTI LoniMover™ is, without a doubt, the easiest and most accurate way to measure CTDI and waveforms for Wide Beam and Cone Beam CT. With the CT Dose Profiler, you will get the FWHM, [...]

Read more

CTDI Phantom Set

Adapt the size of the phantom RTI CTDI Phantom Sets simulates adult and pediatric head and body of patients for QA testing of ­computed tomography (CT) systems. RTI offers two different CTDI Phantom sets. Choose two (2) or three (3) parts. Both have nested modules to allow adaptation to the size required by the selected [...]

Read more

Piranha Panoramic Holder/Vertical Holder v.4

Attach your Piranha vertically The Piranha Panoramic Holder/Vertical Holder v.4 is designed for attaching the Piranha DENTAL meter to dental panoramic equipment or vertical mounting. It can be mounted with an adjustable strap or magnets. [...]

Read more

RTI Light Probe

Designed to comply with the needs of QA in modern X-ray departments The RTI Light Probe measures the brightness on monitors with the same spectral response as the human eye. With a monitor and a lux adapter, the RTI Light Probe measures the brightness on monitors and film viewing boxes and the ambient light in [...]

Read more

RTI MAS-2

The simple way of non-invasive measurements The RTI MAS-2 probe is a clamp-on probe for the non-invasive measurement of mA (tube current) and mAs. Safe and easy to use. The probe is simply clamped onto the high voltage cable and then ready for measurements. No connection inside the X-ray generator is needed! Together with your [...]

Read more

RTI MAS-1

The simple way of invasive measurements The RTI MAS-1 is an invasive probe for the measurement of mA and mAs. Together with Piranha or the Cobia SENSE, direct reading of mA and mAs, as well as waveform, are obtained. The probe connects to the mAs socket in the X-ray generator and to the multimeter. It [...]

Read more

RTI CT Dose Profiler

A pioneer in measuring CT Dose RTI CT Dose Profiler probe for use with our Piranha, Mako, Cobia FLEX, or Cobia SENSE (with PC com). Used for the evaluation of CT and CBCT systems. The probe fits into standard phantoms used for CTDI measurements. One-shot accuracy The RTI CT Dose Profiler has taken CT quality [...]