Cobia SENSE
For quick and accurate constancy checks
Connect different probes and ion chambers to the Cobia SENSE X-ray meter to enable measurements in a wide variety of situations. USB connectivity makes it possible to use the Cobia SENSE together with our Ocean Next™ software on a laptop, PC, or Tablet.
Included: RTI Dose Probe (calibrated at 70 kV, W/3 mm Al (R1)), internal display, internal rechargeable battery, micro-USB cable, power supply, Ocean Next™ software with Quick license, documentation, Soft-Shell transport case, and Basic online training.
For other configurations, please contact your Area Sales Manager or local RTI distributor.
Standard two-year warranty/calibration. Read about our pre-paid calibration and extended warranty.
$4,988
Cobia SENSE for your specific needs
The Cobia SENSE is dedicated for use with an external detector such as the RTI Dose Probe, RTI Light Probe, RTI CT Ion Chamber*, or external mAs probes. The wide selection of external probes enables big flexibility in the performance of regular constancy checks for most modalities.
Customize to your needs by selecting CTDP, DAP, and more.
*Requires a Chamber Adapter to measure.
Auto recognition of external probes
Just connect your probe and you are ready to measure.
Ocean Next™ software
Our Ocean Next™ software exhibits Cobia’s capabilities and adds benefits that make the workflow more practical and simpler. The software makes it possible to perform remote X-ray QA measurements with your Cobia SENSE. You simply connect to your Cobia, via Bluetooth, or with a USB cable to your PC or Tablet, then measure!
With the addition of Ocean Next™ Professional, the Cobia SENSE truly becomes an indispensable solution for data management, efficiency, time-saving procedures, and compliance.
- Large, rotatable display
- Log/History function
- USB communication
- Can be used with ion chambers
- Automatic recognition of external probes
- User menu in nine (9) different languages (English, Chinese, French, German, Japanese, Norwegian, Russian, Spanish, Swedish, Turkish)
- Long-lasting rechargeable battery
- Always free firmware upgrades
- Up to ten-years warranty
- Two-year calibration cycle
-
Weight (approx.) 280 g Size 140 × 78 × 28 mm³ Power source Int. battery / Ext. power supply Battery life 10 – 20 hours Display Color, 320 x 240 pixel Min. Exp. Time 0.1 ms Sensitivity 0.2 mA @ 50 kV, 50 cm (3mm Al) Warranty Two years Extended Warranty Up to ten years Calibration Cycle Two years Standard Complies with relevant parts of the standards IEC 61674 and IEC 61676 for dosimetric instruments PTB Approval DE-15-M-PTB-0005 -
Size 20 x 45 x 7.4 mm, 0.79" x 1.8" x 0.29" Weight 85 g (3 oz) Cable length 2.0 m (6.6 ft) Backscatter protected Yes Dose 6 nGy - 2 kGy. 660 nR - 250 kR (± 5%) Dose rate 220 nGy/s - 220 mGy/s. 1.5 mR/min - 1.5 kR/min (± 5%) Time 0.33 ms – 9999 s Pulses 3 – 9999 pulses Dose per pulse 1 nGy/pulse - 3 kGy/pulse. 114 nR/pulse - 342 kR/pulse Pulse rate/frequency 1/6 - 260 Hz
RTI Dose Probe
For low dose rate measurements The RTI Dose Probe is an external dose probe designed to perform very low dose rate measurements, for instance, on image intensifiers. Use with Piranha, Cobia Dental, Cobia Flex, or Cobia Sense. (The probe can also be used with the Mako system using a Mako Legacy Module.) The probe is [...]
Read moreRTI Light Probe
Designed to comply with the needs of QA in modern X-ray departments The RTI Light Probe measures the brightness on monitors with the same spectral response as the human eye. With a monitor and a lux adapter, the RTI Light Probe measures the brightness on monitors and film viewing boxes and the ambient light in [...]
Read moreRTI MAS-1
The simple way of invasive measurements The RTI MAS-1 is an invasive probe for the measurement of mA and mAs. Together with Piranha or the Cobia SENSE, direct reading of mA and mAs, as well as waveform, are obtained. The probe connects to the mAs socket in the X-ray generator and to the multimeter. It [...]
Read moreRTI CT Dose Profiler
A pioneer in measuring CT Dose RTI CT Dose Profiler probe for use with our Piranha, Mako, Cobia FLEX, or Cobia SENSE (with PC com). Used for the evaluation of CT and CBCT systems. The probe fits into standard phantoms used for CTDI measurements. One-shot accuracy The RTI CT Dose Profiler has taken CT quality [...]
Read moreRTI CT Ion Chamber
The RTI CT Ion Chamber 10 cm and 30 cm are both cylindrical pencil-shaped ionization chambers. They are intended for CTDI measurements free-in-air or in a [...]
Read moreIon Chamber Magna 1 cc
Excellent energy response The Ion Chamber Magna 1 cc is designed for mammography dose measurement. With an excellent energy response, it can be used for radiographic applications too. It is the ideal choice for two reasons: Evaluations of Magna ionization chambers show a response within ±1% over an extended mammography range of 20-50 kVp (HVL [...]
Read moreDAP Chamber
The big news in a small format The DAP Chamber is the perfect tool for field calibration of orthodontic X-ray equipment. You can measure with just one click – fast, easy, and accurate, ensuring your quality control process runs safely and smoothly. Available in two sizes – 86 x 86 mm and 147 x 147 [...]
Read moreRTI Chamber Adapter
External module for connecting ion chambers The RTI Chamber Adapter is an external module for connecting ion chambers to the Piranha, Cobia FLEX, or Cobia SENSE meters. The adapter upports ion chambers with LEMO triaxial connectors (e.g. RTI CT Ion Chamber, DAP chamber, and Magna 1cc mammography ion chamber). Measuring range: 10 pA – 0.7 [...]
Read more10" Tablet PC
Connect wirelessly! Microsoft Surface Go 3 10″ Tablet PC display for wireless or USB communication with the Piranha, Mako and Cobia meters. The portable tablet comes with a detachable keyboard, Windows 11 Home, installed Ocean Next™ software, built-in Bluetooth, a charger, and all necessary items to make it a user-friendly & versatile [...]
Read moreRTI T20 Dose Probe
Use anywhere in the X-ray field The RTI T20 Dose Probe is a solid-state detector dedicated to measurements on Rad/Fluoro systems when it is crucial that the detector itself does not have any effect on the system output or disturb the X-ray beam. It was developed for measurements of the patient entrance dose (skin dose) [...]
Read more